Introduction: Flexible ureterorenoscopy (fURS) is one of the best solutions for treatment of renal calculi <2cm and for upper urinary tract urothelial carcinoma conservative treatment. An adequate quality of vision is mandatory to help surgeon get better outcomes. No studies have been done, to our knowledge, about what fURS in the market has the best quality of vision. Materials and Methods: Seven different fURS were used to compare the image quality (Lithovue, Olympus V, Olympus V2, Storz Flex XCin White Light and in Clara+Chroma modeWolf Cobra Vision, Olympus P6, and Storx Flex X2). Two standardized grids to evaluate contrast and image definition and three stones of different composition were filmed in four standardized different scenarios. These videos were shown to 103 subjects (51 urologists and 52 nonurologists) who had to evaluate them with a rating scale from 1 (very bad) to 5 (very good). Results: No difference in terms of scores was observed for sex of the participants. Digital (D) ureterorenoscopes were rated better than fiber optics (FOs) ureterorenoscopes. Overall, Flex XC White Light and XC Clara+Chroma image quality resulted steadily better than other fURS (p<0.0001). Olympus V generally provided a vision better than Lithovue. Cobra Vision and Olympus V2 had superimposable values that were significantly lower than Lithovue's ones. Olympus P6 and Storz X2 offered a low quality of vision compared to the others. In the medium simulating bleeding, Olympus V and V2 significantly improved their scores of 12% and 8.1%, contrary to rest of the ureterorenoscopes. Conclusion: D ureterorenoscopes have a better image quality than FO ones. The only disposable ureterorenoscope tested was comparable to the majority of other D ureterorenoscopes. The best image quality was provided by Storz D ureterorenoscopes, being Clara Chroma the favorite Spies Mode, according to literature.

Filiaciones:
Talso, M:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France

 Univ Milan, Dept Urol, Fdn IRCCS Ca Granda, Osped Maggiore Policlin, Milan, Italy

Proietti, S:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France

 Osped San Raffaele, Villa Turro Div, Dept Urol, Milan, Italy

Emiliani, E:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France

 Univ Autonoma Barcelona, Dept Urol, Fdn Puigvert, Barcelona, Spain

Gallioli, A:
 Univ Milan, Dept Urol, Fdn IRCCS Ca Granda, Osped Maggiore Policlin, Milan, Italy

Dragos, L:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France

 Univ Med & Farm Timisoara, Dept Urol, Timisoara, Romania

Orosa, A:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France

Servian, P:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France

Barreiro, A:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France

Giusti, G:
 Osped San Raffaele, Villa Turro Div, Dept Urol, Milan, Italy

Montanari, E:
 Univ Milan, Dept Urol, Fdn IRCCS Ca Granda, Osped Maggiore Policlin, Milan, Italy

Somani, B:
 Univ Hosp Southampton NHS Trust, Dept Urol, Southampton, Hants, England

Traxer, O:
 Sorbonne Univ, Tenon Hosp, AP HP, GRC RENAL LITHIASIS 20, F-75020 Paris, France
ISSN: 08927790
Editorial
MARY ANN LIEBERT, INC, 140 HUGUENOT STREET, 3RD FL, NEW ROCHELLE, NY 10801 USA, Estados Unidos America
Tipo de documento: Article
Volumen: 32 Número: 6
Páginas: 523-528
WOS Id: 000430745500001
ID de PubMed: 29562765
imagen Green Accepted

MÉTRICAS